Option
Description
Calibrates the installed sensors.
Calibration
Shows the status screen, event
log, alarm log, sensor data and
internal I/O data. Has a test for
the keypad and display.
Diagnostics
Status indicators
The indicator light shows the status of the instrument. Refer to
LED color
Description
Green
Flashes during normal operation.
Red
Flashes when one or more alarms are active.
Blue
Flashes when there is a communication problem with the IO9000 module or with a sensor with
logged channels.
Orange Flashes when the internal memory battery is low. Contact factory service immediately.
Operation
Configuration options—Instrument or PC
Damage to the logger and PC can occur if the PC is connected to the incorrect port on the logger. Connect the
PC only to the USB B port on the logger.
Use the keypad on the instrument or a PC with the applicable software to configure the instrument.
This document contains instructions to configure the instrument from the keypad.
To use a PC for configuration, refer to the documentation for the applicable software to configure the
instrument. When the configuration is complete, import the configuration file directly from the PC or
from a USB memory stick. Refer to
instrument. Refer to
Import or export the instrument settings
file.
Configure the general settings
Use the general settings menu to get instrument information, change the display settings, erase data,
set security, export/import from a USB port, restore defaults.
1. Push MENU.
2. Select General Settings.
3. Select an option.
22 English
Table 6 Main menu options (continued)
Option
Review Data
General settings
Table 7 LED status indicators
N O T I C E
Figure 1
on page 7 to connect a PC or USB memory stick to the
Description
Shows the measurement data (the data
log).
Shows the instrument details (e.g.,
serial number). Configures the general
instrument and display settings. Erases
data, sets security, exports/imports from
a USB memory stick, restores defaults.
Table
7.
on page 25 to import the configuration