Note : a function can also be assigned via RS232 using the command <FCT + Function No.>
example : Change of Unit = <FCT6>, Changing resolution: = <FCT7>
9 Adjustment of the instrument
- Beforehand, input the dimension of the standard master in the Preset value (see chap. 3)
- Reset the measurement value by a short press on Clear
- Make a standard measurement (going through the turnaround point)
- Adjust the instrument by pressing Preset Master button (>2s)
- The instrument is adjusted and ready to measure.
9.1 Measure
- Make the measurements going through the turnaround point. The digital display stores and displays the MIN
measured value.
- Before each new measurement, reset the measured value by a short press on