Dentsply Sirona Cavitron Mode D'emploi page 8

Inserts ultrasoniques
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
STEAM STERILIZATION
CYCLES
*Dynamic Air Removal
Gravity Displacement
*The steam sterilization parameters specified above are for Dynamic
Air Removal steam sterilizers (pre-vacuum and steam flush pressure
pulse).
Alternate method
Place non-bagged instruments into the steam sterilizer and run at
the listed cycles.
Note: Instruments sterilized unbagged should be used immediately.
Drying
To dry, use the drying cycle of the sterilizer. Set cycle for 20-30
8
minutes.
STERILIZING
STERILIZING
TEMPERATURE
CYCLE TIME
270°F / 132°C
3 minutes
270°F / 132°C
10 minutes
Storage
To maintain sterility, instruments should remain bagged until ready
for use.
WARRANTY
Cavitron® Ultrasonic Inserts manufactured by Dentsply Sirona are
warranted against defects arising from faulty materials or
workmanship for ninety days from the date of purchase. This
warranty extends to inserts purchased from an authorized Dentsply
Sirona distributor and only to the original purchaser and is valid
only for inserts purchased in the United States (U.S.) and for those
countries without specific warranty requirements. For inserts
purchased outside the U.S. in countries with specific warranty
requirements, those applicable country warranties will apply.
This warranty is subject to the following conditions:
1. Inserts must not be subjected to abuse or improper use or
application.