STEAM STERILIZATION
CYCLES
*Dynamic Air Removal
Gravity Displacement
*The steam sterilization parameters specified above are for Dynamic
Air Removal steam sterilizers (pre-vacuum and steam flush pressure
pulse).
Alternate method
Place non-bagged instruments into the steam sterilizer and run at
the listed cycles.
Note: Instruments sterilized unbagged should be used immediately.
Drying
To dry, use the drying cycle of the sterilizer. Set cycle for 20-30
8
minutes.
STERILIZING
STERILIZING
TEMPERATURE
CYCLE TIME
270°F / 132°C
3 minutes
270°F / 132°C
10 minutes
Storage
To maintain sterility, instruments should remain bagged until ready
for use.