Table Des Matières; Specifications; Ppb (Mdl To 0.999 Ppb) ±0.1 Ppb (1.000 To 1.999 Ppb) ±0.5 Ppb Or 5% Whichever Is; Greater (2 To 1999 Ppb) - Hach A1000 Manuel D'utilisation De Base

Masquer les pouces Voir aussi pour A1000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
Table of contents
Additional information
Additional information is available on the manufacturer's website.

Specifications

Specifications are subject to change without notice.
Performance
Specification
A1000
TOC measurement
TOC method
TOC = Total Carbon (TC) - Total Inorganic Carbon (TIC)
Linear range
0.05 to 1999 ppb as
carbon
Detection limit
0.05 ppb
Quantitation limit
0.17 ppb
Accuracy
0.5 ppb or ±5%
whichever is greater
1
The instrument changes to XP mode automatically when the TOC conductivity and temperature range are within the XP mode ranges shown.
2
Resistivity and accuracy specifications are for ambient temperatures from 15-35 °C (59-95 °F).
Maintenance
on page 24
A1000XP
XP mode: 0.02 to 1999 ppb as
1
carbon
Standard mode: 2.00 to
1999 ppb as carbon
0.02 ppb
0.07 ppb
A1000
Precision
±0.5 ppb
TOC conductivity
0.2 to 18.2 MΩ-cm
range
TOC water
0 to 100 °C (32 to
temperature range
212 °F)
Note: Use a heat
exchanger for samples
more than 50 °C (122 °F).
Display resolution
X.YY (0.00 to
19.99 ppb)
X.Y (20.0 to 199.9 ppb)
X (200 to 1999 ppb)
Calibration stability
Annual calibration or validation recommended
Resistivity measurement
Linear range
2
Resistivity accuracy
2% over full range (compensated)
Display resolution
accuracy
Temperature
compensation
±0.05 ppb (MDL to 0.999 ppb)
±0.1 ppb (1.000 to 1.999 ppb)
±0.5 ppb or 5% whichever is

greater (2 to 1999 ppb)

XP mode: 15.0 to 18.2 MΩ-cm
Standard mode: 0.2 to
18.2 MΩ-cm
XP mode: 18 to 32 °C (65 to
90 °F)
Standard mode: 0 to 100 °C
(32 to 212 °F)
Note: Use a heat exchanger for
samples more than 50 °C (122 °F).
X.YYY (0.000 to 1.999 ppb)
X.YY (2.00 to 19.99 ppb)
X.Y (20.0 to 199.9 ppb)
X (200 to 1999 ppb)
0.01 to 18 MΩ-cm
Three significant figures
±0.5 °C (9 °F)
Compensated to 25 °C (77 °F)
English 3

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

A1000xp

Table des Matières