Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Display shows
F 21
F 23
F 24
F 25
F 26

‹ Power failure

The Multimat
Easy offers a power failure protection
®
feature that allows the currently running program to be
continued after a power failure of less than 60 seconds
as soon as the power returns. You will see error message
F 14 (see "Error messages and troubleshooting").
Note: The firing result must be inspected closely even if
the firing program was only interrupted for a very short
time.
If the power failure lasts more than 60 seconds, the
currently running program will be interrupted.
Note: If the firing chamber has to be opened during the
power outage, you may carefully push the firing plat-
form down manually. This will damage neither the motor
nor the gear mechanism.
Problem
Defective program
Vacuum was not relieved
Pressure
Heat sensor broken
Incorrect polarity of heat sensor
Error messages and troubleshooting 9
Possible cause
Memory error
Ventilation valve defective
Heat sensor or heat sensor wiring
broken
Plus and minus poles were
mixed up!
Potential remedy
Controller will delete the defective program
automatically; Program must be re-entered
Notify service technician
Notify service technician
Notify service technician
When shipping the Multimat
please follow the instructions in the
"Safe transport" section.
Do not reach into the firing chamber
– burn hazard!