InterValve V8 Instructions D'utilisation page 8

Cathéter de valvuloplastie aortique transluminale par ballonnet
Table des Matières

Publicité

Prepping the Device
NOTE: DO NOT REMOVE THE BALLOON PROTECTOR UNTIL AFTER THE AIR PURGING
PROCESS IS COMPLETED.
1. Attach a high pressure 3-way stopcock to the balloon inflation port.
2. Attach a 50cc or 60cc syringe filled with approximately 6cc-10cc of sterile normal
saline and inject 1-2cc's of the fluid into the catheter. Draw back on the syringe to
apply full vacuum. Repeat this procedure 3 or 4 times to insure total air evacuation.
Close the stopcock.
3. Fill the inflation syringe with the desired volume (see Volume Compliance Chart) with
a solution of 1:8, or lower concentration (higher dilution), contrast medium to saline
and reattach the syringe.
4. Remove the balloon protector. If bubbles are seen inside the balloon upon removal of
the balloon protector, reapply the balloon protector and repeat step 2 and 3 above
until the bubbles are removed.
Performing the Procedure
1. First shape the distal portion of the guidewire tip into a broad loop to reduce the risk
of perforation and then advance under fluoroscopic guidance to the desired position
within the left ventricle.
2. Advance the balloon over the wire, through the introducer sheath, and across the
valve maintaining a vacuum during delivery, if possible, in order to maintain a
minimal balloon wrap profile.
3. Align the balloon's central marker bands with the valve annulus.
4. Rapid ventricular pacing can be used to aide in balloon fixation during inflation.
5. Rapidly inflate and deflate the balloon with the 50cc/60cc syringe. If pacing
was elected to be used, terminate after balloon deflation. If the balloon bursts,
withdraw the balloon by removing the balloon and sheath together as a unit, as
described in the Warnings section.
6. Repeat steps 3-5, if needed, until desired results are achieved, allowing sufficient
time for complete hemodynamic recovery between inflations.
position throughout the procedure.
7. Antiplatelet and anticoagulant therapy is recommended throughout the procedure to
avoid blood clot(s) which may result in embolization and vascular occlusion.
8. Remove the catheter while applying a continuous vacuum with the syringe and
pulling back slowly on the catheter while holding the introducer sheath firmly in
place.
9. Apply a minimum of 30 minutes of compression to the insertion site after the system
is removed or place an approved vascular closure device to close the arteriotomy
site.
1204-001 Revision D.1
Maintain wire
Page 7 of 11

Publicité

Table des Matières
loading

Table des Matières