Press one of the memory buttons M1 or M2 and the result is
placed in the selected measurement memory. The selected
measurement memory number appears on the display. The re-
sult is stored in the measurement memory selected when the
device is switched off. If a measurement results in an error, it is
not stored (fig. 7).
In order to switch off the
device, press the blue START/
STOP button. Otherwise the
device will switch itself off
automatically after 3 minutes.
Important: If you wish to
repeat the measurement,
allow the wrist to rest for at least 1 minute.
Automatic repeat pumping function
The inflation pressure for each measurement is automatically
set at approx. 190 mmHg. If this pressure is insufficient or if
the measurement is interrupted, the device continues to pump
at a rate of 40 mmHg until a high enough pressure is reached
so that the measurement can start. If you require a higher
inflation pressure, you can avoid having to repeat pumping by
pressing the blue START/STOP button again shortly after
inflation starts and holding down on it until the desired cuff
pressure is reached. This should be approx. 30 mmHg over the
systolic (upper) value. The pressure can be increased to approx.
280 mmHg.
5. Setting the memory function
5.1 Saving the measured values
The measurement result is stored either in M1 or M2. Select the
memory position after measurement. Each of the two memory
positions can store 30 measurement results and the average
value of the results. Once the number of measurements