Disinfecting - Pari LC PLUS Notice D'utilisation

Masquer les pouces Voir aussi pour LC PLUS:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 101

4.3 Disinfecting

After cleaning, disinfect all of the disassembled parts (only
parts that have been cleaned can be disinfected effectively).
The recommended disinfection procedures are described be-
low. Descriptions of other validated disinfection procedures are
available from the manufacturer or dealer upon request.
The connection tube cannot be cleaned or disinfected. Re-
garding care of the connection tube, follow the instructions
in the corresponding section [see: Care of the connection
tube, page 22].
CAUTION
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
• Remove all parts from the pot or disinfector as soon as
the disinfection process is finished.
• Dry all parts.
In boiling water
• Place all the individual parts in boiling water for at least
5 minutes. Use a clean pot and drinking water.
NOTE! Risk of damage to plastic parts! Plastic will melt if it
comes into contact with the hot base of the pot. Make sure
there is enough water in the pot to prevent the individual
parts from touching the pot base.
• Shake the water out of all of the parts.
– 21 –

Publicité

Table des Matières

Dépannage

loading

Ce manuel est également adapté pour:

Lc starLc plus baby022

Table des Matières