5. Put all non-waterproof cables (all cables except the deployment cable) in a dry container. Do not
put cables in a coil any tighter than 30.5 cm (12 in.) in diameter or cable damage can occur.
6. Before storage, clean the sensors. Refer to the sensor documentation.
7. For long-term storage, remove the internal D-cell battery(s). Refer to
the batteries—HL7 sonde
storage.
Figure 16 Prepare for storage or travel
Prepare for use after long-term storage
After the instrument has been in storage for more than 1 month,
1. Replace the electrolyte in the sensors, if applicable.
2. Calibrate the sensors. Refer to
Rehydrate the sensors
If the sensors become dry, soak them in water before use.
Note: Do not keep a total dissolved gas sensor wet or the sensor membrane will be damaged.

Troubleshooting

Refer to the Hydrolab Operating Software Manual for troubleshooting information. To look at the
Hydrolab Operating Software Manual, start the Hydrolab Operating Software, click Help and select
Hydrolab Operating Software Help.

Replacement parts and accessories

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
on page 9. No additional sensor preparation is necessary for long-term
Calibration
on page 11.
W A R N I N G
Figure 5
on page 8 or
English 23