Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz) magnetic field
IEC 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
The Blood pressure monitor SBM 18 is intended for use in the electromagnetic environment specified below. The customer
or the user of Blood pressure monitor SBM 18 should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
All manuals and user guides at all-guides.com
IEC 60601
Compliance
test level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3A/m
3A/m
Compliance
level
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floor are
covered with synthetic material, the relative humidity should be
at least 30%.
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the Blood pressure monitor SBM 18, inclu-
ding cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
65