When
to
perform
a
control
test
Control
Solution
contains
a
check
that
the
meter
and
the
test
strips
are
working
properly.
Do
a
control
solution
test:
•
Whenever
you
open
a
new
vial
of
test
strips.
•
Whenever
you
want
to
check
if
you
are
testing
correctly.
•
If
you
suspect
the
meter
and
test
strips
are
not
working
properly.
•
If
you
have
had
repeated
unexpected
blood
glucose
results.
•
If
you
drop
or
damage
the
meter.
•
If
you
left
the
test
strip
container
open
or
you
think
the
test
strips
have
been
damaged.
•
If
the
test
strips
were
stored
in
extreme
temperatures
and/or
humidity.
•
Your
test
result
does
not
match
how
you
feel.
NOTE:
•
Only
use
the
Sejoy
CS201
control
solution.
•
Close
the
control
solution
bottle
tightly
after
use.
•
Write
the
date
you
open
the
control
solution
bottle
on
the
bottle
label.
The
control
solution
must
be
discarded
3
months
first
open
date
or
the
expire
date
on
the
vial
label,
whichever
comes
first.
•
Do
not
use
control
solution
that
exceeds
the
expired
date
or
discard
date.
•
Refer
to
the
control
solution
label
for
control
solution
storage
conditions.
•
The
control
solution
can
stain
fabric.
Remove
stains
by
washing
with
soap
and
water.
CAUTION:
Do
Not
swallow
control
solution.
It
is
not
for
human
consumption.
Do
Not
apply
control
solution
to
the
skin
or
eyes
as
it
may
cause
irritation.
Control
Solution
Testing
specific
amount
of
glucose
and
is
used
to
after
th
e