4.
It's forbidden to push the bed with force when the casters are braked, which will cause damage
to the braking system.
5.
The wear of the tire tread of the caster can be detected by visual inspection. Some
accumulations such as thread and spun yarn may wrap around the caster. Remove the bolts
and nuts of the caster, clean up the debris, and check whether the bearing of the caster is
damaged. If the parts are not damaged, the caster can be reassembled and used.
6.
Decision whether to replace entire caster is according to the wear condition. After the caster is
replaced or reinstalled, make sure that the axle bolt and nut are connected reliably, and use
lock washers or lock nuts as much as possible.
7.
For casters equipped with brakes, it is necessary to regularly check whether the brakes work
normally, and check the brakes every day or before each use. For a hospital bed equipped with
multiple brake casters, you can lock only one brake caster at a time and try to push the bed to
check whether the braking performance of each caster is good. If the caster brake fails to
function due to wear or damage to the caster, replace the caster immediately and retest the
brake system.
8. If the mechanism of the brake system is damaged, and the brake needs to be repaired or
replaced, please contact our after-sales or an authorized dealer. Every time the brakes are
replaced, the brake performance of the casters needs to be retested.
9. Side rails maintenance
1.
Please check whether the side rails are installed correctly to prevent the lying patient from
shifting and rolling off.
2.
Please always check the screws of the side rail and the side rail fittings, to prevent loosening
and falling off.
3. Please use the red part of the side rail handle correctly, do not use brute force to pull hard,
which will cause damage to the side rail.