4.1 Push rod malfunction handling
Push rod malfunction handling:
(1) Causes of push rods malfunction:
1.
push rod's abnormal Hall signal;
2.
first time programming;
3.
nurse desk, hand controller, guardrails control panel 6S system malfunction.
(2) Push rod malfunction prompting method: if the M1, M3, M4 malfunctions are detected,
JCH35-JSSK Saikang custom hand controller blue indicator is flashing.
(3) Push rod malfunction troubleshooting:
1.
By pressing corresponding buttons, raise the M1pus rod to the top position or lower it to the
bottom position, so the push rod completes the "rebound", and the malfunction is fixed;
2.
By pressing corresponding buttons, lower the M3, M4 push rods to the bottom position,
make them complete "rebound", and fix the malfunction;
3.
When the M3, M4 rods are completely malfunctioned, Reverse TR/TR buttons are not
valid;
4.
When any push rod is malfunctioned, patient mobilization, cardiac position, TR, emergency
buttons are not valid.
4.2 Test method for unresponsive side rail control panel
Step 1, check if the emergency stop switch is pressed
A. Check: if the emergency stop switch is pressed;
B. Check method: turn the emergency stop switch clockwise to open(it bounces out for a short
distance), see pic 1-1;
C. Define a problem: if the emergency stop switch pops up, it means that it has been pressed; if the
emergency stop switch doesn't pop up, it means that it is already in the open state, and there is no
need to rotate it;
D. Conclusion: after the above three steps are finished, test the guardrail, if there is still no
response, go back to step 2.