2.7 Electronic control system composition
2.7.1 Electronic control system diagram
Operator priority:
(1) Nurse control panel > Hand controller > Side rail control panel;
(2) One priority button can interrupt and terminated the same priority button operation (except hand
controller), when one button is released and another one is still pressed, the operation of the last
one is continued;
(3) High-priority keys (except CPR, TR, and STOP) can interrupt and terminate low-priority keys.
When one button is released and another one is still pressed, the operation of the last one is
continued. When CPR, TR or STOP button is pressed during other button operation, CPR, TR or
STOP operation is executed.
2.8.2 Electronic control parts list
JC motor 200 vertical(manual CPR)
JC knee rest motor 100 horizontal
JC bed platform height motor 150 horizontal
JC 3m power cord (environmental)
Name
JC 5-functions control panel
17
Qty
1
1
2
1
1