Mise en service
9.4
Programmes de prélèvement
9.4.1
Différences entre les types de programme
La vue d' e nsemble suivante montre les différences entre les types de programme Basic, Standard
et Advanced :
Basic:
single program
Program start:
immediate
date & time
Standard:
multiple part-program
Program start:
immediate
date & time
volume
Advanced:
multiple part-program
Program start:
immediate
date & time
volume
external signal
*External signal
32
Sampling program (1)
activation immediate
time-, flow proportional
(CTCV, VTCV, CTVV), *
bottle change on time or
# samples*
bottle synchronisation
multiple bottles
Sampling program part n (1 ... 5)
activation immediate, individual
times, multiple date, interval,
deactivation program part 1
time-, flow proportional
(CTCV, VTCV, CTVV), *
bottle change on time or
# samples*
bottle synchronisation
multiple bottles
Sampling program part n (1 ... 24)
activation immediate, individual
times, multiple date, interval,
deactivation program part 1,
parameter event, external signal
time-, flow proportional (CTCV,
VTCV, CTVV), single sample,
sampling table, external signal
bottle change on time or #
samples, external signal, fieldbus
sampling synchronisation
bottle synchronisation
multiple bottles
Liquiport 2010 CSP44
Program stop
program end
continuous
Program stop
program end
continuous
date & time
Program stop
program end
continuous
date & time
external signal
a0020230-de