VII. TECHNICAL SPECIFICATIONS
Items
Battery
Temperature
sensor
Measuring
distance
Measurement
range
Measuring
accuracy
Clinical
repeatability
Display
resolution
Measurement
Time
Memory
Operation
Environment
Storage and
transportation
condition
Weight
Product Size
Expected
Service Life
Automatic
power-off
Measuring site Forehead
Reference
body site
Operation Mode Adjusted mode
Specifications
DC 3V (2 pcs AA batteries)
Infrared Sensor
3cm - 5cm(1.2in - 2in)
Body temperature mode:
32.0°C ~ 43.0°C (89.6°F ~ 109.4°F)
Surface temperature mode:
0°C ~ 60°C (32°F ~ 140°F)
Room temperature mode:
0°C ~ 40°C (32°F ~ 104°F)
Body temperature mode:
32.0°C ~ 34.9°C (89.6°F ~ 94.8°F) :
±0.3°C (±0.6°F)
35.0°C ~ 42.0°C (95°F ~ 107.6°F) :
±0.2°C (±0.4°F)
42.1°C ~ 43.0°C (107.8°F ~ 109.4°F) :
±0.3°C (±0.6°F)
≤0.2°C (0.4°F)
0.1°C (0.1°F)
1 second
32 sets
Ambient temperature: 10°C ~ 40°C
(50°F ~ 104°F)
Relative humidity: ≤85%RH
Atmospheric Pressure: 700 ~ 1060 hPa
Ambient temperature: -20℃ ~ 55℃
(-4℉ ~ 131℉)
Relative humidity: ≤ 95%RH
Atmospheric Pressure: 700 ~ 1060 hPa
Approx. 80g (Battery excluded)
Approx. 175 x 50 x 52 mm (L x W x H)
5 years
≤ 30 secs
Axillary
www.spengler.fr