9. TROUBLESHOOTING
Error
Possible cause
message
Unable to record
a pulse.
The measured
blood pressure is
outside the meas-
urement range.
There is a pneu-
matic system
error.
An error occurred
during the meas-
urement.
The inflation
pressure is higher
than 300 mmHg.
There is a system
error.
The batteries are
nearly flat.
Solution
Please wait one minute and
repeat the measurement.
Ensure that you do not speak
or move during the meas-
urement.
Repeat the measurement.
Ensure that the cu line is
correctly connected and that
you do not move or speak.
Please wait one minute and
repeat the measurement.
Ensure that you do not speak
or move during the meas-
urement.
Please take another meas-
urement to check whether the
cu can be correctly inflated.
Make sure that neither your
arm nor other heavy objects
are pressing on the line, and
that the line is not bent.
If this error message appears,
please contact Customer
Services.
Insert new batteries into the
device.
10. DISPOSAL
Repairing and disposing of the device
• Do not repair or modify the device yourself. Proper operation
can no longer be guaranteed in this case.
• Do not open the device except for the battery compartment.
Failure to comply will invalidate the warranty.
• Repairs must only be carried out by Customer Services or au-
thorised retailers. Before making a complaint, first check the
batteries and replace them if necessary.
• The device must not be disposed of with household waste.
Dispose of the device at a suitable local collection or
recycling point in your country. Dispose of the device in
accordance with EC Directive – WEEE (Waste Electrical
and Electronic Equipment). Please contact the local authori-
ties responsible for waste disposal if you have any questions
regarding disposal.
Disposal of the batteries
• Used, completely discharged batteries must not be disposed
of with household waste. Dispose of the batteries in specially
designated collection boxes, at recycling points or at electron-
ics retailers. You are legally required to dispose of the batteries
correctly.
• The codes below are printed on batteries containing harmful
substances:
- Pb = battery contains lead
- Cd = battery contains cadmium
- Hg = battery contains mercury
33