Immunity test levels
Test
Electrostatic
Discharge
(IEC 61000-
4-2)
Radiated RF
EM filed
(IEC 61000-
4-3)
Compliance level
Contact Discharge:
±8 kV
Air Discharge: ±2 kV,
±4 kV, ±8 kV, ±15 kV
80-2700 MHz; 1kHz
AM 80 %; 10 V/m
Electromagnetic environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P for 80 MHz to 800 MHz
d = 2.3√P for 800 MHz to 2,7 GHz
where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
- 100 -