and against electric or mechanic surcharge.
The thermic power implied in this podoscope avoid dangerous overheating
for materials and components.
Electric isolation and well-balanced distances between the surfaces avoid
the formation of electric arches. In addition to this, the podoscope has a solid
mechanical structure and an excellent stability on the ground.
It has no sharp edges, therefore there exists no danger of mechanical nature
both for operators and patients.
Each part of the inner circuits are assembled and fixed in such a solid way
so as to stand up to the often adverse transport conditions.
As to indirect shocks, the device is classified in Class II.
As to current leakage it can be classified in class B.
No essential performance is defined for this device.
1.5 Electrical features of the podoscope
The conventional led power supply functions with single-phase current
(220-240 V - 50/60 Hz); the mark is ILC NanoLED.
The power cord - H05VV-F - 2 mt (6.56 ft) long, section 2 x 0.75 bears the
mark IMQ, has a two-pole plug and it is connected to the podoscope through
a connector.
The internal connectors of the power cord satisfy the norm CEI 20-20 (type H05V).
1.6 Components identification
1
ENGLISH
7
6
4
5
18
3
2