(13) HFR (High Frequency Rejection)
Available for ACV, ACA and, Flexible Current Probe.
Add ± 4% to specified accuracy of each function and each range
for 45Hz to 200Hz.
Accuracy is unspecified for > 200Hz.
Cut-off Frequency (-3dB) : 800Hz
(14) Inrush Current
Available for ACA and Flexible Current Probe.
Trigger level : ≧50d.
Add ± 3% to specified accuracy of each function and each range.
(15) Resistance
Range
OL Reading
600.0Ω
660.0Ω
6.000kΩ
6.600kΩ
60.00kΩ
66.00kΩ
600.0kΩ
660.0kΩ
To get more accuracy result, short the test probes to obtain the
offset. The accuracy specification is specified for the result that
the offset is subtracted.
Overload Protection : AC/DC 1000V
(16) Continuity
Built-in buzzer sounds when measured resistance is less than
20Ω and sounds off when measured resistance is more than
200Ω, Between 20Ω to 200Ω the buzzer maybe sound or off
either.
Continuity Indicator : 2.7K Tone Buzzer
Response Time of Buzzer : < 100msec.
Overload Protection : AC/DC 1000V
(17) Diode
Range
OL Reading
1.500V
1.550V
Open Circuit Voltage : Approx. 1.8V
Overload Protection : AC/DC 1000V.
Resolution
Accuracy
0.1Ω
±(0.9% + 5D)
0.001kΩ
0.01kΩ
±(0.9% + 2D)
0.1kΩ
Resolution
Accuracy
0.001V
±(0.9% + 2D)
26