Cleaning and Disinfecting
Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must
precede disinfection. Before use by a different individual: clean, disinfect,
and check product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- Follow safety guidelines on the chemical containers.
AGILITY Component
Back shell and hardware:
Cover:
Notes:
- Remove the foam pad from the cover
prior to cleaning and disinfecting.
Secure the hook and loop fastener
strips at both ends of the cover.
- Turn the cover inside out.
- Zip the cover closed to avoid snags.
Foam pad, foam insert and lumbar
support pad:
Note: Remove the foam pad from the
cover prior to cleaning.
Air insert:
Notes: Remove the air insert from the
cover and foam pad prior to cleaning
and disinfecting. Make sure the valve
is closed.
Warnings: DO NOT allow water
or cleaning solution to enter the air
insert. DO NOT machine wash or dry.
DO NOT autoclave. Rinse thoroughly.
Detergent residue may cause air cells
to stick together.
Care Symbols
Wipe with a damp cloth.
Hand wash.
Dry flat in the shade.
Do not tumble dry.
ROHO AGILITY CARBON Contour Back Support Operation Manual
To Clean:
Hand wash with a sponge, using soap and water.
Rinse with clean water. Wipe dry with a
clean cloth and then allow to air dry
completely.
Machine wash cold with mild detergent, gentle cycle,
or wipe clean with a neutral detergent
and water. Rinse thoroughly. Tumble
dry low or air dry.
Wipe lightly with a damp cloth. DO NOT use soap. DO
NOT submerge in water. Wipe off with a clean cloth.
Completely air dry before replacing in
cover.
Close valves. Place in a large sink. Hand wash,
using mild liquid hand soap, hand dishwashing liquid,
laundry detergent, or a multi-purpose
detergent (Follow product label
instructions). Gently scrub all
cushion surfaces, using a soft,
plastic bristle brush, a sponge, or a
cloth. Rinse with clean water. Air dry.
Caution: DO NOT expose to
direct sunlight.
Machine wash warm, normal.
40C
Hand wash warm.
Bleach (1 part bleach: 9 parts water).
1:9
Do not machine wash.
Cautions:
- DO NOT use abrasives such as steel wool or scouring pads on the product.
- DO NOT use cleaning products containing petroleum or organic solvents [i.e.
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids,
adhesive removers] on the product.
To Disinfect:
Wipe gently with a cloth dampened with a household
disinfectant. Rinse with clean water.
Wipe dry with a clean cloth and then
allow to air dry completely.
Hand wash using quaternary ammonium disinfectant
(Follow product label instructions). Rinse thoroughly with
clean water. The cover may also be
40C
machine washed in hot (60°C) water to
disinfect. Tumble dry low. DO NOT use
bleach.
Note: The method of disinfection should be
determined based on disinfection needs.
Note: Disinfectants are NOT effective on porous
surfaces such as foam. Replace soiled or contaminated
foam components with authorized foam components
only. Replacement foam components may be
purchased separately. Soiled or contaminated
foam components MUST NOT be used by different
individuals.
Follow the cleaning instructions, using 1 part household
liquid bleach per 9 parts water. Keep the product wet with
bleach solution for 10 minutes. Rinse thoroughly with clean
water. Air dry.
60C
16
+
Machine wash hot, normal, at the temperature shown.
Tumble dry low.
Do not bleach.
or
60C