OHAUS Pioneer PX Série Manuel D'instruction page 20

Masquer les pouces Voir aussi pour Pioneer PX Série:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

EN-18
Set the following parameters by pressing No or Back:
Water Temperature
Weight
Oil Density
Measure the actual water temperature using a precision
thermometer. The balance calculates water density based on
the water temperature value entered.
Note: The weight of the sample and the density of oil must be
measured in advance.
1. Weigh the oiled sample in the air with the balance and the
density determination kit.
2. When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the oiled
sample in the air.
3. Weigh the oiled sample in the liquid with the balance and
the density determination kit.
4. When the * (symbol of stability) appears, press the
Function / Mode button to confirm the weight of the oiled
sample in the liquid. The density of the sample is displayed.
5. After completion of the test, press the Function / Mode
button to test a new sample.
PX Series Balance

Publicité

Table des Matières

Dépannage

loading

Table des Matières