Patient population
The visionPRO video laryngoscope can be used for all patients that a suitable blade is available for.
Attaching and removing the proctetive case
To attach the optional protective case, attach the protective case to the top front of the DS1 display.
Ensure that the opening for the display hinge is correctly orientated.
Carefully pull the display protection case downwards and pull it over the display frame of the
visionPRO DS1.
To remove the proctective case, proceed in reverse order.
The video laryngoscope's charge level indicator
The charge level indicator is shown by a battery icon in the top part of the display. A distinction is
made between the following charge level indicators:
Battery bar
Solid green
Solid yellow
Solid red
Flashing with a red
exclamation mark
Troubleshooting
If the visionPRO video laryngoscope is not charging, check that the mains plug is connected to
the mains and that the visionPRO video laryngoscope is correctly inserted. If necessary, reset the
charging process by removing the visionPRO video laryngoscope from the visionPRO CS1 for a
few seconds, unplugging the visionPRO CS1 and plugging it back in, and inserting the visionPRO
video laryngoscope. If this error persists, check the electrical contacts for soiling and, if necessary,
clean them carefully with a clean and lint-free cleaning cloth.
If you do not notice any improvement, please contact the manufacturer.
16 | 136
med 235340 2024-08-10
Meaning
The battery bar is green
when the remaining battery
capacity is between 21 %
and 100 %.
The battery bar is yellow
when the remaining battery
capacity is between 11 %
and 20 %.
The battery bar is red
when the remaining battery
capacity is between 6 %
and 10 %.
The battery bar flashes
when the remaining battery
capacity is between 5 %
and 0 %.
Action
No compulsory action is
necessary.
The device should be charged
promptly.
The device must be charged.
The device must be charged.
Further intubation is not
possible.