Using the product
For reliable measurements it is necessary to use reflective
markers. To apply the reflective marker, wait until the object to
be measured has come to a standstill.
The ideal distance to the measurement object is between 10
and 40 cm. During the measurement, always aim at the
reflective marker from an angle of approx.30°. This avoids
multiple reflections from the surface.
If no reflective marker can be attached, it must be checked
whether the reflective properties of a reflective point on the
measurement object are sufficient for rpm measurement. In
this case, aim at the measurement object from the vertical, and
check whether a value is displayed. If the measurement object
displays several reflections per revolution, (e.g. reflective
ventilator with 5 vanes), the number of the impulses per
revolution (e.g. number of ventilator vanes) must be entered
into the instrument (see below) in order to obtain correct rpm
measurements.
Switching t t he i i nstrument o o n:
Press
.
- Measuring mode is opened.
Switching t t he d d isplay l l ight o o n:
The instrument is switched on.
Press
.
- The display light goes out automatically if no key is
pressed for 10 seconds.
Changing t t he d d isplay v v iew/setting t t he n n umber o o f b b lades:
As an alternative to entering the impulses per revolution,
the setting NO.1 can be selected and the rpm calculated
manually from the measurement value (measurement value
divided by impulse/number of vanes)
Adjustable views/setting the number of blades
· Hold: Readings are held.
· Max: Maximum values since the start of the last measurement.
· Min: Minimum values since the start of the last measurement.
· NO.: Set the number of blades. Change the value with
Press
several times until the desired view appears.