1. On the main screen of the motor handpiece, press the
button for 2 seconds. The linkage settings appear.
[ ]
2. Press the
button until the desired menu appears, then
[
]
press the
/
3.
[Auto
start&stop]: Set whether to operate the motor
depending on the location of the file.
• [On]: The motor operates when the file is inserted into the
root canal, and the motor stops when the file is pulled out
from the root canal.
• [Off]: The motor does not run even if the file is inserted
into the root canal. Regardless of where the file is, the
[Main]
button must be pressed to operate the motor.
4.
[Apical
action]: When the Apex Connect measurement
approaches the APEX reference value during root canal
instrumentation, the speed of motor and the torque of the
motor decrease so that the root canal does not instrument
deeper than the set value.
[Speed
slowdown]: When the Apex Connect
measurement approaches the APEX reference value,
the motor's rotational speed decreases.
[Torque
slowdown]: When the Apex Connect
measurement approaches the APEX reference value,
the motor's torque limit decreases.
• [Off]: The motor operates the same regardless of whether
or not the Apex Connect measurement approaches the
APEX reference value.
[Main]
5. Press the
to the main screen.
5
Using Linkage
1. Refer to "USING THE DEVICE > Basic functions" in the
Elements Connect Endodontic Motor Instructions for use to
set the memory number and the motor operation method.
2. Place the lip hook on the lip opposite to the tooth being
treated
3. Slowly insert the Ni-Ti file into the root canal. The root canal
length is measured, and the Apex Connect measurement is
displayed in the
[EAL]
NOTE:
To measure root canal length only,
be set to [Off]. For more information on how to set the motor
operation, refer to "Setting the Linkage" in this manual.
[Main]
4. When the
the motor will start, and the operation is displayed in the
motor screen.
• The following information is displayed in the motor
operation screen:
7
button to select a value.
button. The settings are saved, and return
area of the main screen.
[Auto start&stop]
button is pressed on the motor handpiece,
[ ]
No.
1
2
3
• Confirm the following information with the Apex Connect Unit.
• If the instrument reading approaches the value of the APEX
reference value, an alarm will beep rapidly.
• If the reading reaches or passes the value of the APEX
reference value, the alarm will beep continuously, and the
screen will blink.
• Confirm the measured value range with the lamps of the
[Main]
• If the reading reaches or passes the value of the APEX
reference value, the
will blink.
Lamp Color
• To stop the motor operation, press the
memory number and motor operation settings are changed
on the main screen.
• If selected [CW], [CCW], [Adaptive], or [RCP]as the motor
operation method, press and hold the
main screen to rotate the motor in the opposite direction of
the current setting.
• The motor rotates in the opposite direction only while the
button is pressed and stops when release it.
• Here, the operation screen does not appear on the display,
and the main screen remains on.
5. To end use, press any two buttons on the motor handpiece
at the same time for 2 seconds, then press and hold the
button on the Apex Connect Unit. The power and the display
of both are turned off.
6. To remove pull the lip hook connector from the probe cord.
7. Pull out the USB connectors at both ends of the probe cord
from the motor handpiece and Apex Connect Unit.
8. Refer to "INITIAL SET-UP > Disassembling the Device"
in the Elements Connect Instructions for use to detach the
Ni-Ti file and contra angle from the motor handpiece.
9. Refer to "Cleaning, Disinfection and Sterilization" from
must
the Kerr Endodontic Motor instructions for use to manage
the device and accessories.
PRECAUTION
Be sure to clean and sterilize the lip hook and contra angle.
MAINTENANCE
1
Replacing the Battery
Replace the battery with a new one if there is a problem, such
as when the battery has reached the end of its life or not
charging properly.
1. Remove the silicone ball from the screw hole in the battery
cover on the back of the device. Remove the screw with a
screwdriver, and then remove the battery cover.
Description
• These values represent the APEX reference value and the
measurements taken by the Apex Connect at the file' s
current location. The measured value is also displayed
numerically on the right side of the screen.
• When the measured value reaches the APEX reference
value, the motor operates as follows.
• [R&S]: The motor stops after rotating in the opposite
direction of the current setting.
• [REV]: The motor rotates in the direction opposite to the
current setting. It then rotates in the set direction again.
It is the torque value of the motor when operating and the
torque limit value set in the memory.
[APEX]
blinks when the measurement reaches or passes the
APEX reference value.
button on the motor handpiece.
[Main]
button of the motor handpiece
Lamp Status
Blue
On
Yellow
On
Red
On
Measured value
range
(mm)
3.0–1.1
1.0–0.1
0.0– -0.5
[Main]
button. The
[Main]
button on the
[ ]