Symbols
Symbol
Description
This equipment meets the requirements of all
relevant European safety directives. The equipment
carries the CE mark.
This equipment meets the requirements of all
relevant UK Statutory Instruments. The equipment
carries the UKCA mark.
This symbol, on the equipment, indicates that the
user should read the user manual.
This symbol, on the equipment, indicates a warning
and that the user should refer to the user manual.
Druck is an active participant in the UK and EU
Waste Electrical and Electronic Equipment (WEEE)
take-back initiative (UK SI 2013/3113, EU directive
2012/19/EU).
The equipment that you bought has required the
extraction and use of natural resources for its
production. It may contain hazardous substances
that could impact health and the environment.
In order to avoid the dissemination of those
substances in our environment and to diminish the
pressure on the natural resources, we encourage
you to use the appropriate take-back systems.
Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to
use those systems.
If you need more information on the collection,
reuse, and recycling systems, please contact your
local or regional waste administration.
Please visit the link below for take-back instructions
and more information about this initiative.
https://druck.com/weee
Product Markings
Refer to table and key below:
Product
DPI620G-IS
PM620-IS
PM620T-IS
1. ATEX Certificate Number.
2. UKEX Certificate Number.
3. EU Directive 2014/34/EU Markings.
4. Hazardous Area Markings.
5. IECEx Certificate Number.
6. CE Mark and Notified Body Number (####).
7. UKCA Mark and Approved Body Number (####).
8. Specific Apparatus Type.
9. Serial Number.
Copyright 2016 Baker Hughes Company.
2 | DPI620G-IS Safety Manual–English
Figure
Figure C1
Figure C2
Figure C3
10. Date of Manufacture (Month-Year).
11. Manufacturers Name and Address.
Installation
WARNING
Do not use tools on the
pressure sensor that might cause incendive
sparks - this can cause an explosion.
Only fluids that are compatible with Stainless
Steel and Hastelloy shall be used with the
pressure sensors. This is to ensure the
integrity of the pressure sensor and avoid
fluid leakage.
These instructions detail the requirements for using the
DPI620G-IS in a hazardous area. Read the whole
publication before starting.
•
Installation should be carried out by qualified plant
installation technicians in compliance with all local
safety procedures and installation standards. For
example: IEC/EN 60079-14, US National Electrical
Code NFPA 70 or Canadian Electrical Code (CEC).
•
Provide additional protection for indicators that may be
damaged in service.
DPI620G-IS
Specific Conditions of Use
1. The DPI620G-IS USB Client Connection must only be
connected to external apparatus within a Safe Area
with U
= 254 V.
m
2. When a remote sensor is powered using the Channel
2, 15V, U
loop supply, the remote sensor must be
o
disconnected from all other sources of power.
3. The Channel 2, V terminals can be connected to a
FISCO System if the power for the system is provided
from the DPI620G-IS Channel 2, 15V, U
and the electrical parameters of the field devices are
compatible with those of the DPI620G-IS.
4. The DPI620G-IS should not be connected to a
powered FISCO system unless its defined electrical
parameters are compatible with the DPI620G-IS.
5. If the Channel 1 Terminals and the Channel 2
Terminals are used together then they must always be
connected as separate Intrinsically Safe Circuits.
6. The rechargeable battery pack may be removed from
or replaced in a DPI620G-IS within a hazardous area.
7. The rechargeable battery pack must be removed from
the DPI620G-IS for recharging in the safe area using
only the Socket 3 charging contacts and the Druck
charger with U
= 254 V. No connections must be
m
made to the IS outputs.
8. Both the MC620-IS Dual Transducer Carrier, or the
PV62X-IS Series Pressure Station have a location for
either one or two transducers which screw into the
pressure connection and the two slip rings on the lower
face of a transducer make contact with two spring
loaded pins. When either a MC620-IS Dual Transducer
Carrier, or the PV62X-IS Series Pressure Station is
connected to a DPI620G-IS Advanced Modular
Calibrator, these pins are energised and do not meet
the requirements of at least IP20. Therefore it is a
requirement of certification that before either the
MC620-IS Dual Transducer Carrier or the PV62X-IS
Series Pressure Station is connected to the
DPI620G-IS Advanced Modular Calibrator, a
PM620-IS Transducer or a metallic Dummy Pressure
loop supply
o