Medela Freestyle Mode D'emploi page 9

Masquer les pouces Voir aussi pour Freestyle:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Electrical/ Battery operation 
The pump can be mains adapter and battery operated.
Mains symbol on the display
Battery indicator on the display
4 bar-battery indicator on the display
1 bar-battery  indicator  flashes
Battery  indicator  flashes
"bAtt" flashes on the display
Caution: 
The battery used in this device may present a risk of fire or chemical burn if mistreated.
Do not disassemble, heat or incinerate. Replace battery with Medela part number 200.1512
only. Use of another battery may present a risk of fire or explosion.
When storing the battery for a limited length of time, store within the following temperature
range: -20° to 50° C (-4° to 140° F). When storing the units for an extended time, store
within the following temperature range 0° to 25° C (32° to 77°F). Dispose of used battery
promptly. Keep away from children. Do not disassemble and do not dispose of in fire.
Note
If the pump has been running for 30 minutes without any manipulation, the Freestyle will
automatically switch off.
Avoid removing battery. Removing the battery will reset the battery indicator and requires
a recharging time of up to 24 hours to readjust the battery indicator.
The pump is plugged in.
Battery power state
Full battery, 3h pumping time
20 min.pumping time only
Battery has been removed
Empty battery
Charging for first time
Removal of battery
(Up to 24h for charging-time)
9

Publicité

Table des Matières
loading

Table des Matières