Deleting statistical data
1
You can delete the statistical data (such as the measurement number (NUM), the average
(AVG), the minimum values (MIN) and the maximum values (MAX)).
•
Press and hold the Clear button
data are reset to zero. Measuring can start again.
13. Calibrating the measuring device
•
Press and hold the confirmation button
calibration
•
The calibration mode is set as outlined above in section 'e) Setting calibration mode' in the
section 'Menu settings'. Zero calibration and two-point calibration are available.
Carrying out zero calibration
1
•
If you have selected zero calibration
•
If the following figure is shown on the LC display (2), place the sensor (8) directly on the
uncoated calibration disc. Wait for approx. 2 seconds before you lift the measuring device
off again.
1
•
After lifting, a zero value is displayed. Calibration is complete.
Carrying out two-point calibration
1
•
If you have selected two-point calibration
the non-metallic layer thickness standard (of known thickness) over it.
•
If the following image is shown on the LC display (2), place the sensor (8) directly on the
calibration disc with the standard layer thickness plate. Wait for approx. 2 seconds before
you lift the measuring device off again.
46
(5) for two seconds to delete all saved data. All statistical
or two-point calibration
(4) to switch to the desired calibration mode (zero
).
, use an uncoated calibration disc for calibration.
2
, use an uncoated calibration disc and place