Télécharger Imprimer la page

KERN MPC 300K-1M Mode D'emploi page 65

Masquer les pouces Voir aussi pour MPC 300K-1M:

Publicité

Les langues disponibles

Les langues disponibles

12 Data output RS 232
You can print weighing data automatically via the RS 232 interface or manually by
pressing
via the interface according to the setting in the menu.
This data exchange is asynchronous using ASCII - Code.
The following conditions must be met to provide successful communication between
the weighing balance and the printer.
• Use a suitable cable to connect the weighing balance to the interface of the
printer. Faultless operation requires an adequate KERN interface cable.
• Communication parameters (baud rate, bits and parity) of balance and printer
must match.
12.1 Pin allocation of balance output bushing:
12.2 Technical data
Connection
9 pin d-subminiature bushing
Pin 2 output
Pin 3 input
Pin 5 signal earth
Baud rate
Optional 600/1200/2400/4800/9600
Parity
8 bits,
MPC-BAPS-e-1711
Pin 2: TXD - Output
Pin 3: RXD - Input
Pin 5: GND - Signal ground
27