4.2.4
Using a combination of vacuum and traditional electrodes
Use channel A for the stimulated vacuum electrodes and use channel B
for the traditional electrodes.
1. Connect two vacuum
electrodes to channel A and
short-circuit channel B. See
§4.2.2.
2. Prepare the traditional
electrodes. See the User
Manual of the ET device.
3. Connect the two-ply cable to
the output connector
the Vaco 200.
4.3

The water reservoir is full

1. The LED of the water reservoir
treatment. After you turn the vacuum off, you can not restart the
vacuum.
2. Empty the water reservoir with the drain hose. The start of a
vacuum treatment is possible.
3. Clean the water reservoir. See §5.2.4.
4.4

Reference

4.4.1
Indications
The Vaco 200 can be used for the pathologies from the indication list of
the electrotherapy device. See the User Manual of the ET device.
4.4.2
Contra indications
The contra indications from the electrotherapy device. See the User
Manual of the ET device.
Internal infections
Hemorrhagic risk in the part of the body where the electrodes are to
be placed.
Vaco 200
of
B
light up. Finish the present
16