Guidance and manufacturer's declaration - electromagnetic immunity
The 200-series devices are intended for use in the electromagnetic
environment specified below. The customer or the user of a 200-series device
should assure that it is used in such an environment.
Immunity
IEC 60601
test
test level
Conducted RF
3 V
rms
IEC 61000-4-
kHz 80%
6
150 kHz to 80
MHz
Radiated RF
3 V/m AM 1
IEC 61000-4-
kHz 80%
3
80 MHz to 2,5
GHz
Radiated RF
3 V/m CW 200
ENV 50204
Hz d.c. 50%
895 MHz to
905 MHz
Vaco 200
Compliance level Electromagnetic environment -
AM 1
10 V......0,15-80
Mhz
26 V...........6,78
Mhz
32 V.........13,56
Mhz
14 V.........27,12
Mhz
27 V.........40,68
Mhz
10 V/m..0,08-1,0
Ghz
26 V/m....1,4-2,0
Ghz
30 V/m...433,92
Mhz
30 V/m........915
Mhz
30 V/m.895-905
Mhz
191
guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of a 200-series device,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
d = 0,35÷ p
d = 0,46÷ p
d = 0,38÷ p
d = 0,86÷ p
d = 0,44÷ p
80 MHz to 800
d = 0,35
p
MHz
800 MHz to 2,5
d = 0,70÷ p
GHz
d = 0,12÷ p
d = 0,23÷ p
d = 0,23
p