I
NSTALLATION
2.1
Receipt
1. Check whether the equipment has been damaged during transport.
2. Check whether the accessories are intact and complete. See §7.4.
•
Inform your supplier of any damage or defects by no later than
within 3 working days after receipt. Report the damage by
telephone, fax, e-mail or letter.
•
Do not use the equipment if it is damaged or defective.
2.2
Placing and connection
1. Place the equipment on a horizontal and stable base.
•
Do not place the equipment in the sun or above a heat source.
•
Do not use the equipment in a wet area.
2. Check whether the voltage that is stated on the rear of the
equipment corresponds with the voltage of your mains supply. The
equipment is suited for a nominal mains voltage from 100 V to 240
VAC / 50-60 Hz.
3. Connect the device to an outlet with protective earth terminal.
2.3
Use in combination with another device
The Vaco 200 can be used in combination with:
•
The Combi200.
•
The Combin200L.
•
The Duo200.
2.4
Transport and storage
Take account of the following matters if the equipment has to be
transported or stored:
•
Transport or store the equipment in the original packaging.
•
The maximum period for transport or storage is: 15 weeks.
•
Temperature: -20 °C to +60 °C.
•
Relative humidity: 10% to 100%.
•
Atmospheric pressure: 200 hPa to 1060 hPa.
2.5
Reselling
This medical equipment must be traceable. The equipment and some
accessories have a unique serial number. Provide the dealer with the
name and address of the new owner.
Vaco 200