5.2.4
Cleaning the vacuum hoses and the water reservoir
1. Empty the water reservoir with the drain hose.
2. Connect the vacuum hoses.
3. Put the ends of the vacuum hoses in a 70% alcohol solution.
4. Turn the vacuum unit on.
5. Suck the liquid up until the LED of the water reservoir lits.
6. Turn the vacuum unit off.
7. Empty the water reservoir.
8. Repeat steps 3 until 7 with pure water.