Télécharger Imprimer la page

Bühler technologies RC 1 Serie Notice De Montage page 20

Compresseur refroidisseur de gaz de mesure
Masquer les pouces Voir aussi pour RC 1 Serie:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
RC 1.x
Fuse is trig-
• Increased com-
gering
pressor current
consumption
due to faulty
compressor
start-up
Modbus com-
• Bus connection
munication
fault
error
• Line termination
fault
• Check bus con-
figuration
7.1.1 Error Messages in the Display
If an error occurs, the display will read "Err". Press the "
button to show the error number(s).
Error messages will appear until the unit has been restarted
or the error is cleared using the "Func" button. It can only be
cleared if the cause for the error has been corrected.
Causes / Action: The following is a list of the most common
causes and actions for the respective error. If the actions listed
do not resolve the problem, please contact Service.
Problem /
Possible cause
malfunction
No display
• No voltage
• Loose connect-
ing cable
• Display defect-
ive
D1.02 (The software ver-
sion for the display
(perman-
will appear).
ent)
• Not commu-
nicating with
the controller
Error
• An error has
occurred
Error
• Controller
01
malfunction
Error
• Microcontrol-
03
ler fault /
MCP2
Error
• EEPROM error
04
Error
• Moisture de-
22
tector 1 cable
break
Error
• Moisture de-
32
tector 2 cable
break
20 Bühler Technologies GmbH
• Compressor PTC not
cooled down suffi-
ciently. Wait 5
minutes and try
again.
• Check electrical con-
nections
• Check bus line
• Check/reset config-
urator
"
Action
• Check the supply
cable
• Check fuse
• Check connec-
tions
• Check connec-
tions
• Read the error
number as de-
scribed above
• Clear error (tem-
porary fault)
• Disconnect from
power for ap-
prox. 5 s
• Contact service
• Contact service
• Contact service
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
Error
• General error
40
temperature
sensor 1
Error
• Low temperat-
41
ure / short-cir-
cuit temperat-
ure sensor 1
Error
• Excess temper-
42
ature / short-
circuit temper-
ature sensor 1
Error
• Measurement
43
fluctuation
temperature
sensor 1
Status text
Possible cause
H2o.1
• Moisture
alarm mois-
ture detector 1
H2o.2
• Moisture
alarm mois-
ture detector 2
init
• Initialisation
phase
PuMP
• Pumps deac-
tivated
• Excess/low
temperature
(Flashing)
8 Disposal
The refrigerant circuit of the cooler contains R134a refrigerant.
The heat exchanger is charged with glycol-based coolant.
The applicable national laws must be observed when dispos-
ing of the products. Disposal must not result in a danger to
health and environment.
The crossed out wheelie bin symbol on Bühler Technolo-
gies GmbH electrical and electronic products indicates special
disposal notices within the European Union (EU).
The crossed out wheelie bin symbol indicates
the electric and electronic products bearing
the symbol must be disposed of separate
from household waste. They must be properly
disposed of as waste electrical and electronic
equipment.
Bühler Technologies GmbH will gladly dis-
pose of your device bearing this mark. Please
send your device to the address below for this
purpose.
We are obligated by law to protect our employees from haz-
ards posed by contaminated devices. Therefore please under-
stand that we can only dispose of your waste equipment if the
device is free from any aggressive, corrosive or other operat-
ing fluids dangerous to health or environment. Please com-
plete the "RMA Form and Decontamination Statement", avail-
• Sensor possibly
defective
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
• Check temperat-
ure sensor con-
nection
Action
• Dry
• Check condens-
ate trap
• Dry
• Check condens-
ate trap
• Wait
• Reactive pumps
via menu
• see chapter
"Troubleshoot-
ing"
BX450028 ◦ 03/2024

Publicité

loading

Ce manuel est également adapté pour:

Rc 1.2+ serieRc 1.1 serie