•  Do not use the Change Indicator system on residents who experience
frequent fecal incontinence. TENA SmartCare Change Indicator does
not detect fecal episodes.
•  Keep the Change Indicator and Change Indicator batteries away from
children and pets.
•  Seek immediate medical care if the battery is swallowed.
•  If a defibrillator has been used on a resident who was wearing the
Change Indicator, the Change Indicator must be replaced.
•  Do not swing the Change Indicator in the direction of people, pets or
nearby objects. Be aware of the risk of strangulation with the sensor strip.
•  Clean the Change Indicator when fully assembled to prevent water
from entering the device. Do not submerge in water or clean in a
washing machine. Do not put the Change Indicator in an autoclave.
•  Ensure adequate disinfection procedures are followed for residents
with infectious diseases.
•  If a resident is experiencing loose bowel movements or diarrhea,
wait until the condition has been treated before resuming use of the
Change Indicator.
•  The Change Indicator and gateway should not be used in an
oxygen-rich environment.
•  The Change Indicator and gateway should not be used in the presence
of flammable anesthetics.
•  The Change Indicator is not intended for use in an aseptic
environment.
•  The Change Indicator is not magnetic resonance safe and should
not be brought into a magnetic resonance imaging environment.
•  Do not use the Change Indicator where X-Rays are performed.
•  Do not attempt to repair or modify the Change Indicator or a gateway
by yourself.
•  If any part of the Change Indicator system appears to be damaged
or broken, it should not be used. Contact TENA SmartCare support
in your country.
•  Do not expose the battery to high temperatures, disassemble it,
damage it mechanically or put battery into the fire as this can cause
an explosion or a fire.
TENA SmartCare Change Indicator
5
TM/MC

WARNINGS