Recirculation Cooling Unit - Struers Discotom-2 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Subject
Electrical Data
Mains Cable
Recommendation
Residual Current
Circuit Breaker
Environment

Recirculation cooling unit

Voltage / frequency:
Motor Output
Pump Capacity
Rate of Protection
Dimensions and Weight
Autofeed
16
Discotom-2
Instruction Manual
Specification
Voltage / frequency:
3X220V 50HZ
13.2 A
3X220V 60HZ
7.6 A
3X380V 50HZ
6.9 A
3X415V 50HZ
7.6 A
3X440V 60HZ
6.0 A
3X480V 60HZ
5.5 A
Min.
Fuse
size
3X220V 50HZ
16 A
3X220V 60HZ
15 A
3X380V 50HZ
10 A
3X415V 50HZ
10 A
3X440V 60HZ
10 A
3X480V 60HZ
10 A
Important:
Local standards may overrule the recommendations for the main supply cable.
If necessary, please contact a qualified electrician to verify which option is suitable for the
local installation setup.
type A, 30 mA (or higher) is recommended.
Noise level
Max. 80 dB (A) at a distance of 1 m from the machine
Safety standards
Please refer to the Declaration of Conformity
3 x 110-120 V/50-60 Hz
3 x 220-415 V/50-60 Hz
3 x 420-480 V/50-60 Hz
3 x 500-580 V/50-60 Hz
140 W
57 l/min at 1 m water column
IP 44
Height
Width
Depth
Weight
Air Pressure:
Max. consumption:
Minimum cable size
Max.
@ Min. fuse
Fuse size
3x1,5mm² + PE
25 A
3xAWG14 + PE
25 A
3x1,0mm² + PE
25 A
3x1,0mm² + PE
25 A
3xAWG16 + PE
25 A
3xAWG16 + PE
25 A
Metric
410 mm
400 mm
500 mm
7 kg
Min. 4 bar, max. 9.9 bar
approx. 5l/min
Minimum cable size
@ Max. fuse
3x1,5mm² + PE
3xAWG14 + PE
3x1,0mm² + PE
3x1,0mm² + PE
3xAWG14 + PE
3xAWG14 + PE
Imperial
16"
16"
20"
15 lbs