KIT COMPONENTS
Unpack and inspect the components for possible damage in shipment. If any damage is
noted, immediately notify the carrier and supplier and do not use the light source.
EagleEye™ EK-3000 Kit
9
3
4
5
1.
EagleEye™ EE-365 UV-A/White Light LED Lamp
2.
Splash Guard with Integral Particulate Filter (2), SG-100
3.
Battery Charging Cradle and AC Cord Set (100V–240V)*, 128217
4.
Lithium Ion Batteries (2), 149828
5.
12V DC Cord Set, 128225
6.
Lamp Mount/Sprayer, LMS-100
7.
Head Strap, HS-100
8.
UV-Absorbing Spectacles, UVS-30
9.
Soft Carrying Case, CC-370A
*Each of the AC cords is fitted with the appropriate plug for the country of destination.
IMPORTANT
You must wear the UV-absorbing spectacles supplied. The lamp should be used in
low-light conditions to achieve the best inspection results.
USING THE LAMP MOUNT/SPRAYER
•
To attach the lamp, pull back the metal clip at the top of the
sprayer, slide the lamp onto the mounting bracket and
release the clip.
•
Remove the cap from the penetrant can. Place the jaws
of the sprayer above and behind the opening on the can.
Squeeze the tabs to attach the sprayer under the first ridge
on the can. Do not touch the sprayer's lever until you are
ready to spray.
8
6
7
1
2
B
Sprayer's Lever
Tab
A
4
2
REPLACING THE SPLASH GUARD
The EagleEye EE-365 lamp comes equipped with two splash guards that are also available
in replacement packs. They protect the LEDs from penetrant. In addition, one side of the
guard is fitted with a particulate filter that protects the internal fan from foreign matter.
To replace, hold the lamp firmly in one hand and gently pull the splash guard from the top
with the fingers of the other hand. Align a new one with the indicators on top next to the white
LEDs and press into place.
Splash Guard
with Integral
Particulate Filter
REPLACING THE UV LENSES
•
The UV lenses that cover the two UV-A LEDs must be replaced periodically, since they
2
will grow cloudy (solarize) over time from repeated use.
•
Remove the splash guard and set it aside.
•
Remove the four Phillips screws in the faceplate and retain them. Remove and set aside
the faceplate.
•
Replace the solarized lenses with new ones. Reattach the faceplate with the four
screws.
•
Replace the splash guard.
ATTACHING THE LAMP TO THE HEAD STRAP
Adjust the head strap to fit your head or helmet. The adjusting pin on the strap's mounting
bracket has been factory-tightened to hold the lamp firmly in position. Use an Allen wrench to
adjust the pin further, if necessary. To attach the lamp to the strap, pull the mounting bracket
away from the strap bracket. Pull back the metal clip on the mounting bracket and slide the
lamp down onto the bracket. Release the metal clip to secure the lamp. Place the strap and
lamp on the helmet, securing the strap under its back lip. Position the lamp by moving it
forward and back until the lights are focused on the work surface.
B
1
A
UV INTENSITY VERIFICATION
To ensure that your EagleEye EE-365 lamp is operating at the required UV intensity, it
should be checked periodically. The AccuMAX
B
PROPERTY OF SPECTRONICS CORP; DO NOT DUPLICATE
a dual-wavelength UV-A/VIS sensor detector to accurately measure ultraviolet irradiance
or visible light. The AccuMAX XR-1000 digital readout unit with the XS-365 UV sensor or
the Spectroline DM-365XA digital readout unit are also recommended for accurate UV-A
measurement. The meters are specially designed to measure the UV irradiance from
320–400nm with a peak at 365nm.
For more information about these units, please contact the Customer Service Department at
Spectro-UV. Call toll-free 1-866-230-7305.
A
1
UV-A Lens
Faceplate
4
3
Metal Clip
Mounting
Strap
Bracket
Bracket
Adjusting Pin
FILE NAME & PATH
MATERIAL
DIMENSIONAL TOLERANCES
UNLESS OTHERWISE SPECIFIED
FRAC.
+
1/64
-
SCALE
TM
XRP-3000 radiometer/photometer kit uses
NEXT ASSY.
4
3
5
2
1
REVISIONS
SYM.
DESCRIPTION
DATE
APPROVAL
B
N:\Standards\Template2010\Spectronics\SStandard.idw
A
FINISH
SPECTRONICS PART NO.
TITLE
SPECTRONICS
DEC.
ANG.
CORPORATION
+
+
.005
30'
-
-
956 BRUSH HOLLOW ROAD
WESTBURY, NEW YORK 11590
NO. REQ'D
ChrisF
12/7/2010
DRAWN
DATE
SIZE
DRAWING NUMBER
REV
CHECKED
DATE
B
APPROVED
DATE
2
1