3. Installation / 4. Differential pressure gauge
Option
Panel mounting
GB
4. Differential pressure gauge
The measuring range of the differential pressure gauge can, depending on the particular
measurement system used, be adjusted within the limits given in the previous specifications
table). Ideally, this adjustment should be made on a test bench, though it can also be carried
out at the measuring point using a hand test pump.
Measuring range limits
Measuring cell
140 mbar
280 mbar
560 mbar
1130 mbar
2300 mbar
4000 mbar
Adjustable measuring span
Turn clockwise: reduce measuring range
Turn anticlockwise: expand measuring range
WIKA operating instructions differential pressure gauge models 712.15.160, 732.15.160
6
Adjustable measuring ranges
from
to
0 ... 40 mbar - 0 ... 140 mbar
0 ... 80 mbar - 0 ... 280 mbar
0 ... 160 mbar - 0 ... 560 mbar
0 ... 320 mbar - 0 ... 1130 mbar
0 ... 650 mbar - 0 ... 2300 mbar
0 ... 1150 mbar - 0 ... 4000 mbar
Adjustable measuring span
Allen key
(included in delivery)
Panel cut-out
Panel
Cover cap for adjustable
measuring span