Portable radio communications equipment (such as WIFI
router, smart watches or mobile phones etc.) may interfere
with the performance of the Lenire device (including cables
as specied by Neuromod Devices Limited) and should
be kept at least 2m away from it when in use. Otherwise,
degradation of the performance of this equipment could result.
If interference to the performance of the device occurs, such
as a change in the display indicators, gaps or drop-outs in
the sound, or a noticeable and unusual change in the tongue
stimulation, it is possible that a source of electromagnetic
interference is in proximity to the device. Please move to
a location that signicantly reduces the proximity to the
electromagnetic interference source.
The Lenire device should not be used adjacent to or stacked
with other electrical equipment because it could result in
improper operation.
The Headphones may suer damage if exposed to
Electrostatic Discharges of higher then 8 kV. To minimise
the risk of damage, please pick up the Headphones by the
headband before use.
Use of accessories and cables other than those specied
for use for the Lenire device (and as provided by Neuromod
Devices Limited) could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
The Lenire device should not be used while in proximity to
diathermy, electrocautery, magnetic resonance imaging,
security or metal detector systems. Otherwise, degradation
of the performance of this equipment could result.
Lenire has been tested for electromagnetic emissions and complies
with Class B limits of CISPR 11 (Controller and Tonguetip) and CISPR 32
(Headphones) and is therefore unlikely to interfere with the operation
of other electrical equipment.
Chapter 26 - Materials in Direct Contact with Human Tissue
The Lenire device does not contain phthalates, latex, human blood
derivatives or cells or tissues of human or animal origin.
For a complete list of materials in contact with human tissue, contact
Neuromod Devices Limited.
IFU-0026 Lenire User Manual Version 5.0
Chapter 27 - Declaration of Conformity Radio Equipment Directive
2014/53/EU
Hereby Neuromod Devices Limited declares that the radio equipment
type are Wideband Data Transmission Equipment in the 2.4GHz ISM
band (operating in the frequency range 2,4 GHz to 2,4835 GHz per
ETSI 300 328 v.2.1.1.) is in compliance with Directive 2014/53/EU. The
full text of the EU declaration of conformity is available at the following
internet address: www.neuromoddevices.com.
Chapter 28 - Headphone Specic Technical Information
Use these Headphones only for the intended applications. Austrian
Audio cannot accept any liability for damages resulting from improper
handling or misuse.
Complies with IEC 62368-1:2018 for mobile audio equipment.
Hereby, Austrian Audio GmbH declares that the radio
equipment type Hi-X25BT is in compliance with Directive
2014/53/EU. The full text of the EU declaration of conformity
is available at the following internet address:
https://austrian.audio
Headphone Manufacturer Contact Details:
Austrian Audio GmbH
Eitnergasse 15, 1230 Vienna, Austria
The Bluetooth® word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc. and any use
of such marks by Austrian Audio GmbH is under license.
USB Type-C® and USB-C® are registered trademarks
of USB Implementers Forum.
Page 17