10. Microphotography
10.1 Use of C-mount adapter
1. Loosen the clamping screw ① on the trinocular port and re-
move the dust cap ②. (Fig. 35)
2. Screw the C-mount adapter ③ to the camera ④ and insert
the round dovetail of the C-mount into the empty hole of the
trinocular port (Fig. 36), then tighten the clamping screw ①.
10.2 Use of Reflex cameras
1. Insert the Reflex adapter ① into the relay tube ②.
2. Screw the "T2" ring ③ (not provided) to the reflex adapter.
3. Connect the Reflex camera ④ to the "T2" ring just installed.
(Fig. 37)
4. Mount the other end of the relay tube ② into the empty hole
of the trinocular port, then tighten the clamping screw. (Fig.
35)
•
"T2" ring is not provided along with the microscope, but is
commercially available.
•
While shooting dark specimens, darken eyepieces and view-
finder with a dark cloth to minimize the diffused light.
•
To calculate the magnification of the camera: objective mag-
nification * camera magnification * lens magnification.
•
When using an SLR camera, mirror movement may
cause the camera to vibrate.
•
We suggest lifting the mirror, using long exposure times
and a remote cord.
Page 20
④
③
①
②
①
F ig. 35
F
ig. 35
④
③
F ig. 36
F
ig. 36
②
F ig. 37
F
ig. 37