Name and model
Measurement range
Laboratory measurement
precision
Interval between
2 measurements
Clinical repeat precision
Measurement units
Ambient operating
conditions
Storage conditions
Dimensions
Weight
Battery
Memory
Signs and symbols
FT 65
Ear/Forehead thermometer mode: 93.2 °F – 109.4 °F (34°C – 43°C)
Object temperature mode: 32 °F – 212 °F (0°C – 100°C)
Ear thermometer mode: ±0.4 °F (±0.2°C) from 95.9 °F – 107.6 °F (35.5°C – 42°C),
outside this measurement range ±0.5 °F (±0.3 °C)
Forehead thermometer mode: ±0.5 °F (±0.3 °) from 93,2 °F – 109,4 °F (34°C – 43°C)
Object temperature mode: ±2.7°F (±1.5°C) at < 86 °F (30°C); ±5% at ≥ 86 °F (30 °C)
At least 5 seconds
Ear: Children, 1 – 5 years: ±0.14 °F (±0.08 °C)
Adults: ±0.13 °F (±0.07 °C)
Forehead: Children, 1 – 5 years: ±0.13 °F (±0.07 °C)
Adults: ±0.14 °F (±0.08 °C)
°Celsius (°C) or °Fahrenheit (°F)
60.8 °F – 95 °F (16°C – 35°C) with a relative humidity of up to 85%
(non-condensing)
-13 °F – 131 °F (-25°C – 55°C) with a relative humidity of up to 85%
(non-condensing)
38.2 x 138 x 46.5 mm
90 g including battery
2 x 1.5V AAA (LR03)
For 10 measurements
Device classifi cation type BF
Follow instructions for use
Keep dry