Table 4
Recommended separation distances between portable and mobile RF communications equipment and the EQuIPMENT or
SYSTEM-for EQuIPMENT and SYSTEMS that are not LIFE-SuPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the FT 65 Multi-function
Thermometer
The FT 65 Multi-function Thermometer is intended for use in an electromagnetic environment in which radiated RF distances are con-
trolled. The customer or the user of the T FT 65 Multi-function Thermometer can help prevent electromagnetic interference by maintain-
ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the FT 65 Multi-function Ther-
mometer as recommended below according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
112
Separation distance according to frequency of transmitter
150 khz to 80 Mhz
d = 1.2
0.12
0.38
1.2
3.8
12
(m)
80 Mhz to 800 Mhz
d = 1.2
0.12
0.38
1.2
3.8
12
800 Mhz to 2.5 Ghz
d = 2.3
0.23
0.73
2.3
7.3
23