Perform Calibration opens the calibration menu. Make sure you have run the self-test before performing any
calibration. Refer to "Calibration Adjustment Procedures" in the Service Guide for details.
Press Change Passcode to change the passcode using the numeric keypad. The passcode can be set up to 9 digits.
Auto save On will automatically save the calibration data upon logout. When Auto save is Off, press Cal Save to save
the calibration data.
Self-Test
Self Test verifies proper instrument operation.
Self-Test Procedures
A power-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the
power supply is operational.
Press Utilities > Test / Setup > Self Test to perform the complete self-test of the power supply. It takes approximately
2 seconds for the self-test to complete.
You can also perform a complete self-test from the remote interface, see the E36200 Series Programming Guide for
details.
– If the self-test is successful, "Self test passed" is displayed on the front panel.
– If the self-test fails, "!Err" is displayed on the front panel. Record the error code and message and contact Keysight
support if necessary.
– If the self-test is successful, this indicates a high chance that the power supply is operational.
User Settings
User Settings specifies user preferences that control how the user interacts with the instrument. These settings are
stored in non-volatile memory.
Sound Settings
Beeper enables or disables the beeper sound when an error is generated from the front panel or remote interface.
Key Click enables or disables the click heard when a front-panel key or softkey is pressed.
Keysight E36200 Series User's Guide
117