Connect the 115/230V/24 Vdc power adapter to the CTV101 "POWER
IN" socket then connect the power adapter to a mains socket (115 Vac
or 230 Vac / 50 or 60 Hz) using the mains cord supplied.
For safety reasons make sure that the power adapter is placed at a
distance of at least 1 metre from a water supply.
Important
If the EDI101 electrode is used in a corrosive atmosphere, the electrode
gas inlet (4) (see previous page) must be connected to an inert gas (i.e.
nitrogen) bottle. If this is not done, the electrode bearings and contacts
may quickly be damaged by corrosion and/or clogging agents.
While using the electrode for the first time, the electrical contact must be
run in at 3000 rpm for 20 minutes (see chapter Use to adjust the rotation
speed). Avoid removing socket (3) which holds this contact without any
reasons. The contact must be run in each time you remove socket (3).
9