Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Rear panel

Rear panel
Rear panel
Rear panel
Rear panel
Socket for external control (10)
This BNC socket is to be connected to an external DC voltage generator
when using the external control mode of the instrument (3-position
switch (9) of the CTV101 front panel turned to "EXTERNAL CONTROL").
.
central contact: input of the generator signal
(accepted signal: DC voltage between 0 and 5 V),
.
shielding: connected to the electrical zero of the instrument.
Socket for EDI101 (11)
This 5-pin DIN socket is to be connected to the EDI101 Rotating Disc
Electrode using the cable supplied with the instrument.
Pin layout:
1
3
5
4
2
The functional ground socket (12)
Non insulated socket. Can be used to shield the electrochemical cell
using a Faraday's cage for example (in this case, this socket is to be
connected to the Faraday's cage).
Power socket (13)
This socket is to be connected to a mains socket via the
115/230V/24 Vdc power adapter supplied with the instrument.
Warning: We decline any responsibility in the event of use of an adapter
other than the one provided by Radiometer Analytical.
Pins 1 and 3: motor power supply
(1 = "+" pole, 3 = "-" pole)
Pin 2: electrical ground
Pins 4 and 5: sensor (power supply and pulse output)
7
xxxRyyyNzzz
xxxRyyyNzzz
MADE IN FRANCE
MADE IN FRANCE
RADIOMETER ANALYTICAL SAS
RADIOMETER ANALYTICAL SAS
EDI101
EDI101
POWER IN
POWER IN
EXT
EXT
24 Vdc 500 mA
24 Vdc 500 mA
-
+
0-5 V
0-5 V

Publicité