Aperçu de l'installation – installation dans le bâtiment / branchement fixe
Building instalation
ATTENTION
Control cabinet
(Eco)RO Dia I/II C
Rév. 2.5 Date 2017-03-20
Version logicielle 2.0
ATTENTION
Follow the nationals requirements and of the
local network operator for the installation
provided by customer.
Pre-fuse
(Eco)RO Dia I C => max. 16AT
(Eco)RO Dia II C => max. 16AT
FI-circuit breaker
(Eco)RO Dia I C => 25A/30mA/4pole
(Eco)RO Dia II C => 25A/30mA/4pole
Device/circuit breaker
(Eco)RO Dia II C => min. 16A
(Eco)RO Dia I C => min. 16A
The power switch must not break
the grounding conductor!
The power switch must be placed
in the near of the device and for
the user easily accessible
The separator must meet demands
IEC 60947-1 and IEC 60947-3.
The function of the power switch
as separator of the RO from the
power line must be labeled.
Line filter
Netzfilter
Main switch
Hauptschalter
L1 L2 L3
N
PE
-F
T1
T2
T3
N
PE
-Q
L1
L2
L3
N
PE
Ferrit
-1Z1
-1V1
1
3
5
N
L1
L2
L3
-1Q1
2
4
6
N
T1
T2
T3
Plan de mise en place et schéma des bornes
Partie 2 – Compléments aux instructions d'utilisation
(Eco)RO Dia I/II C
PE
PA-Montageplatte
PA-Mounting plate
Page 8-3