PERFORMANCE DATA AND
ORDERING INFORMATION
Volume
Division
Inaccuracy* (E%)
µL
µL
Min.
vol.
0.2-2
0.1
<± 6.0%
1)
<± 4.0%
0.05
<± 2.5%
<± 1.5%
1-10
1-10
0.05
<± 2.5%
<± 1.5%
2-20
0.1
<± 2.5%
<± 1.5%
10-100
0.5
<± 1.5%
<± 0.9%
20-200
1.0
<± 1.5%
<± 0.9%
100-1000
5.0
<± 1.5%
<± 0.6%
320.336
Workstation for 6 single channel pipettes
320.337
Workstation for 7 single channel pipettes
320.340
Workstation for 3 single, macro or multi-channel pipettes
* Values obtained with forward pipetting mode, at constant tem-
perature (+/–0,5°C) comprised between 20 and 25°C with Aqua
bidest according to ISO 8655. Viscous or volatile liquids may
generate deviations greater than shown in above figures.
1)
At 0.5 µL.
SPARE PARTS
4
TECHNISCHE DATEN UND
BESTELLINFORMATION
Imprecision* (CV%)
Mid.
Max.
Min.
Mid.
vol.
vol.
vol.
vol.
<± 2.0%
<5.0%
1)
<3.3%
<± 1.0%
<2.0%
<1.3%
<± 1.0%
<2.5%
<1.6%
<± 1.0%
<1.7%
<1.1%
<± 0.8%
<1.0%
<0.6%
<± 0.6%
<0.6%
<0.4%
<± 0.5%
<± 0.5%
< 0.4%
* Diese Werte wurden mit Normal Pipettiervorgang, bei einer
konstanten Raumtemperatur (+/–0.5°C) zwischen 20 und 25°C
mit destilliertem Wasser erreicht, gemäss ISO 8655. Hochvisko-
se oder entspannte Flüssigkeiten können zu Resultaten führen,
die den oben aufgeführten Daten nicht genau entsprechen.
1)
Mit 0.5 µL.
ERSATZEILE
822.0002
822.0010
DESCRIPTION
BESCHREIBUNG
2. Circlip
2. Halbmondring
10. Plunger spring
10. Feder
16. Pad
16. Polster
18. Lock pin
18. Stift
21. Clip
21. Clip
22. Small cam
22. Kleine Lehre
25. Large cam
25. Grosse Lehre
27. Ejector bar
27. Abwerfstange
29. Overshoot spring
29. Überhubfeder
32. O-ring
32. O-ring
40. Detachable nozzle
40. Abnehmbarer Konus
53. Bracket
53. Klammer
54. Circlip
54. Circlip
55. Overshoot stop
55. Überhubanschlag
56. Catch
56. Sperrklinke
57. Guide
57. Führungshülse
58. Plunger assembly
58. Kolbenaggregat
59. Lower stop
59. Unterer Anschlag
60. Plunger stop
60. Kolbenanschlag
61. Rack, large cam
61. Zahnring, grosse Lehre
62. Rack, small cam
62. Zahnring, kleine Lehre
63. Ratchet
63. Klinke
64. Spacer
64. Distanzhülse
65. Spring plate
65. Feder-Lamelle
66. Handpiece
66. Handgriff
67. Window
67. Scheibe
68. Colour plate
68. Farbplakette
69. Ejector button
69. Abwurfknopf
70. Setting wheel
70. Einstellrad
75. Calibration screw
75. Kalibrierschraube
76. Push button
76. Druckknopf
78. Screw
78. Schraube
79. Ejector ring
79. Drücker
82. Barrel assembly
82. Zylinderaggregat
83. Spacer
83. Distanzhülse
SOCOREX ISBA S.A. – CH. CHAMP-COLOMB 7 – CH-1024 ECUBLENS/LAUSANNE – SWITZERLAND
socorex@socorex.com – www.socorex.com – TEL. +41 21 651 6000 – FAX +41 21 651 6001
Socorex
Selection of Qualitips
Cat. No.
Description
Max.
vol.
<1.5%
822.0002
Nat.
<0.5%
822.0010
ultra-microtip
Nat. gel loading
<0.7%
822.0010Y
Yellow and
<0.5%
822.0020
nat. tips
<0.2%
822.0100
<0.2%
822.0200
Nat. Gel loading
< 0.2%
822.1000
Blue and nat.
tips
B = bag, F = filter, L = long, R = rack, S = single wrapped, AF = autocl. filter,
G = gel loading
822.0010Y
822.0020
822.0100
822.0200
DESCRIPTION
2. Circlip
10. Ressort
16. Coussinet
18. Goupille
21. Clip
22. Petite came
25. Grande came
27.Tige d'éjecteur
29. Ressort de surcourse
32. O-Ring
40. Pointe détachable
53. Attache
54. Circlip
55. Butée de surcourse
56. Cliquet
57. Guide butée
58. Assemblage piston
59. Butée inférieure
60. Butée mobile
61. Crémaillère, grande came
62. Crémaillère, petite came
63. Cliquet
64. Entretoise
65. Lamelle ressort
66. Poignée
67. Vitre
68. Plaquette de couleur
69. Bouton éjecteur
70. Molette de réglage
75. Vis d'étalonnage
76. Bouton-poussoir
78. Vis
79. Poussoir d'éjecteur
82. Assemblage cylindre
83. Entretoise
DONNÉES TECHNIQUES ET
CATALOGUE
Pipette tips. Ask for detailed leaflet.
®
Volume
Cat.No.
Volume
µL
Without filter
µL
10
309.0010R + B
10
10
302.0010GR
200
307.0200R + B
20
308.0200R + B
100
328.0200R + B
200
308.0350R + B
300
308.0200GR
100
1000
307.1000R + B
1000
318.1000R + B
319.1000R + B
* Valeurs obtenues en mode de pipettage direct, avec de l'eau
distillée à une température ambiante constante (+/–0.5°C) com-
prise entre 20 et 25°C selon la norme ISO 8655. Le pipettage de
solutions visqueuses et hautement volatiles peut conduire à
des déviations plus importantes que celles mentionnées ci-
dessus.
1)
A 0.5 µL.
PIÈCES DE RECHANGE
822.1000
DECLARATION OF CONFORMITY
SOCOREX ISBA S.A.
CHAMP-COLOMB 7 - 1024 ECUBLENS/LAUSANNE
SWITZERLAND
We declare under our sole responsability that our devices
ACURA
are in conformity with the following Council directive
IVD 98/79/EEC
(in vitro diagnostic medical devices)
Sylvain Christen
Director, CEO
Ecublens
May 10 , 2 0
th
10
Cat.No.
With filter
309.0010FR
308.0020FR + FS
308.0100FR
308.0200LFR + LFS
308.0300FR + AFB
308.1000FR + FS + AFB
309.1000FR + FS + AFB
manual pipettes
®
Jean-Pierre Uldry
Director, Engineering