Operate the device in a way that it is not difficult to operate the
network separators.
Always adjust the rotary switch to the desired measuring range
before starting the measurement and engage the switch in the
proper measuring range.
Never turn the rotary switch during measurement, but always in the
disconnected condition.
Never connect the device to voltage or current sources that exceed
the specified maximum values.
Disconnect the power supply and discharge the filter capacitors in
the power supply before measuring resistance or testing diodes.
Never connect the test leads of the device to a voltage source, if the
rotary switch is set to measure current, resistance or test diodes. This
can damage the device.
If the battery symbol appears in the display, replace the battery
immediately.
Always switch off the appliance and remove the test leads from all
voltage sources before opening the device to exchange the battery or
the fuse.
Never use the device with the rear cover removed or with the battery
and fuse compartment open!
Do not use the device near strong magnetic fields (for e.g. welding
transformer), as this can distort the display.
Do not use the device outdoors, in humid surroundings or in
environments that are subjected to extreme temperature
fluctuations.
Do not store the device in places which are exposed to direct
sunlight.
Remove the battery if the device is not used for a long time.
If changes or modifications are made to the device, the operational
safety is no longer guaranteed and the warranty becomes void.