Commands should be followed by the ASCII code <CR>
?
Demand of displayed value
FNC x
Select measuring function:
X=
0
1
Func=
—
A
FNC?
Demand of measurement function
ID?
Demand of identification of instruments TE TT20 or TE TT60
MM / IN
Change of working unit
MES?
Type of measurement, internal or external (INT, EXT)
PRE xxx
Input offset value of the present function
xxx = real (x.xxxx or -x.xxxx for mm / x.xxxxxx or -x.xxxxxx for inch)
PRZ
Offset value = 0
RNG x
Select analogue scale
X=
0
1
Range µm
±5000
±2000
Range inch
±.200
±.100
RNG?
Transmits used scale
RST
Reset: puts instrument into default configuration
STO0, STO1 Deactivate/ activate, measurement mode
TOL xxx xxx Input of tolerance limits
xxx = real (x.xxxx or -x.xxxx for mm / x.xxxxxx or -x.xxxxxx for inch)
TOL?
Instrument transmits the value of tolerance limits
TO0, TO1
Deactivate/ activate, measurement mode with tolerance
UNI?
Demand of measuring unit mm or in
VER?
Demand of the program version
ONLY FOR TT60
MEM x
Select the mode: Min, Max, Max-Min and (Max+Min)/2
X =
0
1
Func. MEM= Normal Max
MEM?
Demand of the present memory function
STM
Initiate the memories to the present value
2
3
4
-A
B
-B
2
3
4
±500
±200
±50
±.020
±.010
±.002
2
3
4
Min
Max-Min (Max+Min)/2
5
6
7
A+B
A-B
-A+B
5
6
7
±20
±5
Auto
±.001
±.0002
Range
8
-A-B