• Main cleaning and maintenance operations

seat and remove manifold 4 (Fig. C) pulling the filter outwards
by its handle. To check filter 10 lift bent pipe union 11 (Fig.
γ
D). For Aspi-Jet 6
, every evening, once disconnected the
power supply line, loosen the two rubber bands and remove
lid: take out canister, empty and clean (Fig. E).
γ
For Aspi-Jet 7-8-9
the canister must be removed fortnight,
cleaned and disinfected.
Canisters, covers and probes of all models once a week
must be accurately cleaned with a sponge and Puli-Jet
plus diluted in hot water.
Every evening, after cleaning the filters, it is important to
aspirate Puli-Jet plus diluted in hot water; to prepare the
Puli-Jet plus solution follow the instructions on the bottle.
Disinfection and cleaning should be carried out by means of
Pulse Cleaner (Fig. F). Once rinsing is completed, Aspi-Jet
γ
7-8-9
carry out drainage automatically, while for Aspi-Jet
γ
6
the canister is to be emptied once again.
Puli-Jet plus New with anti-scale dissolves blood and
mucus and performs an antimicrobial action; if used
regularly, every day with hot water, Puli-Jet plus removes
old scalings and disagreable smells.
Do not use detergents, even with reduced foaming, as
aspirated air volume and turbulence may cause foam build
up and damage the suction unit, make it stop and produce
disagreable smells.
O-rings (tightening rings) and sliding closures of terminals
(Fig. G) must be lubricated with Lubri-Jet spray every 15
working days. It is advisable to replace all hose of the plant
(Fig. H), expecially outside the unit, and terminals periodically
for sanitary and functional reasons (flexibility and lightness
of hose, smoothness of sliding closures).
• Main cleaning and maintenance
operations
• After every patient, replace or disinfect the external hose
(preferably with Eco-Jet 1).
• After every surgical operation or any long operation: rinse
the appliance by aspirating hot water (50 °C).
• At noon, before lunch time, clean the system with Puli-Jet
plus (sanitizing at 4% - disinfecting at 8%).
• After each working day: clean filters, clean plant with
disinfecting Puli-Jet plus and hot water.
• Once every fortnight clean the canister, draining valve and
probes; lubricate OR and sliding closures of terminals with
spray silicon.
Fig. F
Fig. G
Fig. H
21